CVE-2017-12714
moderate-risk
Published 2018-04-25
Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Do I need to act?
-
0.21% chance of exploitation
EPSS score — low exploit probability
-
Not on CISA KEV list
No confirmed active exploitation reported to CISA
?
Patch status unknown
Check vendor advisories for fix availability and mitigation guidance
6
CVSS 6.5/10
Medium
ADJACENT_NETWORK
/ LOW complexity
Affected Products (7)
Accent Firmware
Anthem Firmware
Accent Mri Firmware
Accent St Firmware
Assurity Firmware
Allure Firmware
Assurity Mri Firmware
Affected Vendors
References (4)
Third Party Advisory
http://www.securityfocus.com/bid/100523
Third Party Advisory
https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01
Third Party Advisory
http://www.securityfocus.com/bid/100523
Third Party Advisory
https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01
36
/ 100
moderate-risk
Severity
21/34 · High
Exploitability
1/34 · Minimal
Exposure
14/34 · Moderate